Monte Carlo modeling of a holder for irradiation of dosimeters in beta radiation beams
Carregando...
Data
2019
Data de publicação:
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
ODS
Resumo
A study of the influence of a dosimetric holder and its cover for the irradiation of the dosimeters in the Beta
Secondary Standard BSS2 radiation fields is reported. The correction factors for attenuation or scattering were
calculated taking into account the BSS2 beta source energy, as well as the detector source distance. The study
and the determination of these factors were carried out using the Monte Carlo Method. The results of the correction
factors showed that the absorption and scattering of the electrons depend greatly on the energy of beta
radionuclides. The determined correction factors are in agreement with the international report ISO 6980.
Como referenciar
POLO, IVON O.; SANTOS, WILLIAM S.; CALDAS, LINDA V.E. Monte Carlo modeling of a holder for irradiation of dosimeters in beta radiation beams. Applied Radiation and Isotopes, v. 143, p. 60-66, 2019. DOI: 10.1016/j.apradiso.2018.10.020. Disponível em: http://repositorio.ipen.br/handle/123456789/30005. Acesso em: 19 Apr 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.