X-ray spectrometry applied for determination of linear attenuation coefficient of tissue-equivalent materials

Carregando...
Imagem de Miniatura
Data
2019
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radiation Physics and Chemistry
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
Resin-based materials equivalent to water were developed and four different samples were obtained. The linear attenuation coefficients of all samples was evaluated using X-ray spectrometry with primary and transmitted beams using voltages at the ranging from 60 to 120 kV. The experimental measured values were compared with theoretical reference values to water and with that obtained using the Least Square Method algorithm methodology (method applied to diagnostic radiology and radiotherapy). Our results show that differences between the measured values and the target μ(E) was lower than 7% ± 0.3 in the energy range from 20 to 80 keV. These results enable to consider that the material developed and produced is a good option to be used as a waterequivalent material and the experimental method adequate to its quantitative evaluation.

Como referenciar
FRIMAIO, AUDREW; NASCIMENTO, BRUNA C.; BARRIO, RAMON M.M.; CAMPOS, LETICIA L.; COSTA, PAULO R. X-ray spectrometry applied for determination of linear attenuation coefficient of tissue-equivalent materials. Radiation Physics and Chemistry, v. 160, p. 89-95, 2019. DOI: 10.1016/j.radphyschem.2019.03.021. Disponível em: http://repositorio.ipen.br/handle/123456789/30057. Acesso em: 16 Apr 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções