Evaluation of an extrapolation chamber for dosimetry in computed tomography beams using Monte Carlo code (MCNP5)
Carregando...
Data
2020
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radiation Physics and Chemistry
Resumo
Computed tomography (CT) is responsible for the highest dose values to the patients during the exams.
Therefore, the radiation doses in this procedure must be accurate. However, there is no primary standard system
for this kind of radiation beam yet. A homemade extrapolation ionization chamber was preliminary evaluated
for the establishment of a CT beam primary standard. The aim of this study was to determine the influence that
the various components of the extrapolation chamber may present on the energy deposited in its sensitive
volume. Different materials for its collecting electrode were also studied for this type of radiation beam. The
evaluation of the homemade extrapolation chamber showed the highest influence of 25.9% for the collecting
electrode, and the best material for the collecting electrode for CT radiation beams was graphite.
Como referenciar
CASTRO, MAYSA C.; SILVA, NATALIA F.; SANTOS, LUCAS R.; CINTRA, FELIPE B.; CALDAS, LINDA V.E. Evaluation of an extrapolation chamber for dosimetry in computed tomography beams using Monte Carlo code (MCNP5). Radiation Physics and Chemistry, v. 167, p. 1-4, 2020. DOI: 10.1016/j.radphyschem.2019.04.043. Disponível em: http://repositorio.ipen.br/handle/123456789/31435. Acesso em: 19 Apr 2024.