The response of low-cost photodiodes for dosimetry in electron beam processing

Carregando...
Imagem de Miniatura
Data
2021
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radiation Physics and Chemistry
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The response of thin diodes (SFH206k) as dosimeters has been investigated employing the beam of an electron accelerator within the dose rate range of 2–8 kGy/s and accumulated doses up to 100 kGy. These devices, operating in the short-circuit mode and under industrial irradiation conditions, deliver current signals nonlinearly dependent on the dose rate, whichever the dose history of the diodes, due to the high density of the generated electron-hole pairs herein achieved. Despite this nonlinearity, the dose rate response is stable and characterized by current signals with repeatability better than 2.0%, regardless of the accumulated dose. It is also found that the dose responses are quite linear with sensitivities slightly dependent on the accumulated dose at a constant dose rate. The decrease in the charge sensitivity, taking as reference that obtained before any radiation damage, reaches only 9% (k = 2) at 100 kGy, which is much smaller than the values reported in the literature. From this low aging and the repeatability of both dose rate and dose responses, it seems that the photodiode under investigation is a low budget alternative, good enough for routine dosimetry, provided it has been previously calibrated in the same processing facility.

Como referenciar
GONCALVES, JOSEMARY A.C.; MANGIAROTTI, ALESSIO; ASFORA, VIVIANE K.; KHOURY, HELEN J.; BUENO, CARMEN C. The response of low-cost photodiodes for dosimetry in electron beam processing. Radiation Physics and Chemistry, v. 181, p. 1-8, 2021. DOI: 10.1016/j.radphyschem.2020.109335. Disponível em: http://repositorio.ipen.br/handle/123456789/31961. Acesso em: 19 Apr 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções