Comparison of calibration of surface contamination monitors with Brazilian Network - 2018/2019

Carregando...
Imagem de Miniatura
Data
2021
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Journal of Physics: Conference Series
ODS
ODS 7ODS 9ODS 12
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
This work reports the results obtained from the comparison involving 7 laboratories in Brazil. This exercise is about the calibration service of surface contamination monitors. The monitor has been calibrated to BS ISO 7503-3, the calibration factor in terms of surface emission rate. The comparison was conducted by the Brazilian National Laboratory of Ionizing Radiation Metrology (LNMRI / IRD) from October 2018 to July 2019. The extensive sources used were 14C, 137Cs, 60Co, 90Sr/90Y, 36Cl and 241Am. The result of this proficiency test was excellent and proved the calibration capacity of the Brazilian network in the calibration service for surface contamination monitors.

Como referenciar
CABRAL, T.S.; LARANJEIRA, A.S.; POTIENS, M.P.A.; SOARES, C.M.A.; KHOURY, H.; SAITO, V.; DAVID, M.; VIDAL, L.C.; PIRES, M.A. Comparison of calibration of surface contamination monitors with Brazilian Network - 2018/2019. Journal of Physics: Conference Series, v. 1826, p. 1-6, 2021. DOI: 10.1088/1742-6596/1826/1/012039. Disponível em: http://repositorio.ipen.br/handle/123456789/32032. Acesso em: 23 Apr 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções