Comparison of calibration of surface contamination monitors with Brazilian Network - 2018/2019
Carregando...
Data
2021
Data de publicação:
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Journal of Physics: Conference Series
ODS
Resumo
This work reports the results obtained from the comparison involving 7 laboratories
in Brazil. This exercise is about the calibration service of surface contamination monitors. The
monitor has been calibrated to BS ISO 7503-3, the calibration factor in terms of surface
emission rate. The comparison was conducted by the Brazilian National Laboratory of Ionizing
Radiation Metrology (LNMRI / IRD) from October 2018 to July 2019. The extensive sources
used were 14C, 137Cs, 60Co, 90Sr/90Y, 36Cl and 241Am. The result of this proficiency test was
excellent and proved the calibration capacity of the Brazilian network in the calibration service
for surface contamination monitors.
Como referenciar
CABRAL, T.S.; LARANJEIRA, A.S.; POTIENS, M.P.A.; SOARES, C.M.A.; KHOURY, H.; SAITO, V.; DAVID, M.; VIDAL, L.C.; PIRES, M.A. Comparison of calibration of surface contamination monitors with Brazilian Network - 2018/2019. Journal of Physics: Conference Series, v. 1826, p. 1-6, 2021. DOI: 10.1088/1742-6596/1826/1/012039. Disponível em: http://repositorio.ipen.br/handle/123456789/32032. Acesso em: 23 Apr 2024.