Radiology and mammography standard X-ray spectra simulated with the Monte Carlo method

Carregando...
Imagem de Miniatura
Data
2021
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
Six standard beams described in the TRS-457 (IAEA): RQR 5, 8, M1, M2, M3, M4 were simulated using the EGSnrc Monte Carlo code. Each spectrum was created by an X-ray tube simulated in BEAMnrc, and attenuation curves were obtained using the application egs_kerma. The quality of each beam was evaluated by the 1st and 2nd half-value layers, the homogeneity coefficients and the mean energies. All beams presented quality parameters compatible with those described in TRS-457 (IAEA).

Como referenciar
XAVIER FILHO, JOEL M.; SILVEIRA, I.S.; CALDAS, L.V.E. Radiology and mammography standard X-ray spectra simulated with the Monte Carlo method. Brazilian Journal of Radiation Sciences, v. 9, n. 2C, p. 1-13, 2021. DOI: 10.15392/bjrs.v9i2C.1670. Disponível em: http://repositorio.ipen.br/handle/123456789/32645. Acesso em: 20 Apr 2024.