BACCARELLI, A.M.DIAS, M.S.KOSKINAS, M.F.2014-11-172014-11-182015-04-022014-11-172014-11-182015-04-02BACCARELLI, A.M.; DIAS, M.S.; KOSKINAS, M.F. Standardization of sup(241)Am by means of coincidence system using 4 pi plastic scintillator detector. In: REGIONAL CONGRESS ON RADIATION PROTECTION AND SAFETY, 5th, REGIONAL IRPA CONGRESS, Apr. 29 - May 4, 2001, Recife, PE. <b>Proceedings...</b> DisponÃvel em: http://repositorio.ipen.br/handle/123456789/15241.http://repositorio.ipen.br/handle/123456789/15241A 4nt (a,(3)- y coincidence system using a plastic scintillator detector in 47t geometry has been developed and applied for the standardization of 241Am. The scintillator shape and dimensions have been optimized for maximum charge particle detection efficiency, while keeping background low and a nearly constant gamma ray efficiency for different points at the radioactive source. The gamma ray events were measured with 3" x 3" NaI(Tl) scintillation counter. The electronic system for processing pulses consisted of logic gates and delay modules feeding a Time to Amplitude Converter with output to a Multichannel Analyser. The alpha detection efficiency measured with 241Am resulted around 95% and the beta detection efficiency measured with 60Co was around 60%. Activity measurements of 241Am were compared with a conventional system employing a 4it proportional counter and showed good agreement.openAccessamericium 241standardizationplastic scintillation detectorsnai detectorscoincidence methodsgamma radiationalpha detectionbeta detectioncobalt 60radioactivityStandardization of sup(241)Am by means of coincidence system using 4 pi plastic scintillator detectorTexto completo de eventohttps://orcid.org/0000-0001-8407-9833https://orcid.org/0000-0003-2478-5757