REGINA BECK TICIANELLI

Projetos de Pesquisa
Unidades Organizacionais
Cargo

Resultados de Busca

Agora exibindo 1 - 3 de 3
  • Artigo IPEN-doc 25771
    Evaluation of the uncertainty associated with sample holders in NAA measurements in LAN/IPEN
    2019 - ZAHN, G.S.; TICIANELLI, R.B.; SAIKI, M.; GENEZINI, F.A.
    In IPEN’s Neutron Activation Laboratory (LAN/IPEN), thin stainless steel sample holders are used for gamma spec-trometry in NAA measurements. This material is very practical, but its chemical composition may be troublesome, as it presents large amounts of elements with intermediate atomic number, with attenuation factors for low-energy gamma-rays that must not be neglected. In this study, count rates obtained using different sample holders were compared. To accomplish that, an Am-241 source, with 59-keV gamma emission, was used so that low-energy gamma attenuation differences can be determined. Moreover, in order to study the energy dependence of these differences, a Ho-166m source was also used. From these results, it was possible to analyze the experimental error associated to the variations between sample holders, with the aim of introducing an addictive term to the uncertainty analysis of comparative Neu-tron Activation Analysis results.
  • Artigo IPEN-doc 24074
    Evaluation of the uncertainty associated with sample holders in NAA measurements in LAN/IPEN
    2017 - ZAHN, GUILHERME S.; TICIANELLI, REGINA B.; SAIKI, MITIKO; GENEZINI, FREDERICO A.
    In IPEN’s Neutron Activation Laboratory (LAN/IPEN), thin stainless steel sample holders are used for gamma spectrometry in NAA measurements. This material is very practical, but its chemical composition may be troublesome, as it presents large amounts of elements with intermediate atomic number, with attenuation factors for low-energy gamma-rays that must not be neglected. In this study, count rates obtained using different sample holders were compared. To accomplish that, an Am-241 source, with 59-keV gamma emission, was used so that low-energy gamma attenuation differences can be determined. Moreover, in order to study the energy dependence of these differences, a Ho-166m source was also used. From these results, it was possible to analyze the experimental error associated to the variations between sample holders, with the aim of introducing an addictive term to the uncertainty analysis of comparative Neutron Activation Analysis results.
  • Artigo IPEN-doc 21161
    Uncertainty analysis in comparative NAA applied to geological and biological matrices
    2015 - ZAHN, GUILHERME S.; TICIANELLI, REGINA B.; LANGE, CAMILA N.; FAVARO, DEBORAH I.T.; FIGUEIREDO, ANA M.G.