    Tests with different kind of kVp-meters in standard X-ray beams used for instruments calibration, in diagnostic radiology level
    2008 - LUCENA, RODRIGO F. de; POTIENS, MARIA da P.A.; CALDAS, LINDA V.E.; VIVOLO, VITOR
    In this work, tests were carried out in one X-ray system, Medicor Mövek Röntgengyara, Model NeoDiagnomax, (single –phase, full wave rectified) that operate in the range from 40 kV to 100 kV, and fixed 6 mA in the fluoroscopic mode, with total filtration of 2.5 mmAl. It was used as reference system an invasive system from Radcal, model Dynalyzer IIITM. The equipments tested were three non-invasive instruments, one PTW, model DiavoltTM, and two Victoreen, model NEROTM. It was performed measurements with those equipments, in the range from 50 kV to 100 kV, in steps of 10 kV and current of 6 mA (fixed). The quantity measured was kVpmax. The results showed the maximum variation of 3 % for one NEROTM and 1 % for the other equipment of the same type, and 0,3 % for DiavoltTM related to the invasive reference measurement system. The differences in the obtained values could be caused by the different kind of equipment used in that, their manufacturer calibration and aged.
  • Artigo IPEN-doc 18288
    Evaluation of the voltage quantities measured with different noninvasive meter for quality control at a calibration laboratory
    2012 - VIVOLO, VITOR; NEVES, LUCIO P.; PERINI, ANA P.; SILVA, JONAS O.; LUCENA, RODRIGO F.; POTIENS, MARIA da P.A.; CALDAS, LINDA V.E.