Artigo IPEN-doc 24418 Filtration influence in a constant potential X-ray machine peak voltage measurements2017 - SANTOS, L.R.; VIVOLO, V.; NAVARRO, M.V.T.; XAVIER, M.; POTIENS, M.P.A.This work shows the peak voltage measurements for several beam filtrations used in diagnostic radiology, using two types of non-invasive detectors; a voltage meter and a high-resolution spectrometer. The technique chosen for the voltage peak measurements with the spectrometer was the endpoint. The results were compared to the measured ones and showed good similarity to the nominal values. However the voltage meter detector used in this work pre-sented errors for heavier filtrations.Artigo IPEN-doc 23585 Mammography calibration qualities establishment in a Mo-Mo clinical system2016 - CORREA, E.L.; SANTOS, L.R. dos; VIVOLO, V.; POTIENS, M.P.A.In this study the mammography calibration qualities were established in a clinical mammography system. The objective is to provide the IPEN instruments calibration laboratory with both mammography calibration methods (using a clinical and an industrial system). The results showed a good behavior of mammography equipment, in terms of kVp, PPV and exposure time. The additional filtration of molybdenum is adequate, air-kerma rates were determined and spectra were obtained.Artigo IPEN-doc 21690 Mammography calibration qualities establishment in a Mo-Mo clinical system2015 - CORREA, E.L.; SANTOS, L.R. dos; VIVOLO, V.; POTIENS, M.P.A.