Time response measurements of rosemount pressure transmitters (model 3154) OF Angra I Power Plant
dc.contributor.author | SANTOS, ROBERTO C. dos | |
dc.contributor.author | PEREIRA, IRACI M. | |
dc.contributor.author | JUSTINO, MARCELO C. | |
dc.contributor.author | SILVA, MARCOS C. | |
dc.coverage | Internacional | pt_BR |
dc.creator.evento | INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE | pt_BR |
dc.date.accessioned | 2018-01-03T17:23:33Z | |
dc.date.available | 2018-01-03T17:23:33Z | |
dc.date.evento | October 22-27, 2017 | pt_BR |
dc.description.abstract | This paper shows the Response of time five Rosemount model 3154N pressure trnasmiter from the Angra I Nuclear Power Plant. The tests were performed using the Hydraulic Ramp and Pressure Step Generator from the Sensor Response Time Measurement laboratory of CEN - Nuclear Engineering Center of IPEN. For each transmitter, damping was adjusted so that the time constant was less than or equal to 500 ms. This value has been determined so that the total value of the protection chain response time does not exceed the established maximum value of 2 seconds. For each transmitter ten tests were performed, obtaining mean values of time constant of 499.7 ms, 464.1 ms, 473.8 ms, 484.7 ms and 511.5 ms, with mean deviations 0.85%, 0.24%, 0.97%, 1.26% and 0.64% respectively. | pt_BR |
dc.event.sigla | INAC | pt_BR |
dc.identifier.citation | SANTOS, ROBERTO C. dos; PEREIRA, IRACI M.; JUSTINO, MARCELO C.; SILVA, MARCOS C. Time response measurements of rosemount pressure transmitters (model 3154) OF Angra I Power Plant. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 22-27, 2017, Belo Horizonte, MG. <b>Proceedings...</b> Rio de Janeiro, RJ: Associação Brasileira de Energia Nuclear, 2017. Disponível em: http://repositorio.ipen.br/handle/123456789/28209. | |
dc.identifier.uri | http://repositorio.ipen.br/handle/123456789/28209 | |
dc.local | Rio de Janeiro, RJ | pt_BR |
dc.local.evento | Belo Horizonte, MG | pt_BR |
dc.publisher | Associação Brasileira de Energia Nuclear | pt_BR |
dc.rights | openAccess | pt_BR |
dc.subject | damping | |
dc.subject | hydraulics | |
dc.subject | nuclear power plants | |
dc.subject | performance testing | |
dc.subject | pressure measurement | |
dc.subject | response functions | |
dc.subject | sensors | |
dc.subject | time measurement | |
dc.subject | angra-1 reactor | |
dc.title | Time response measurements of rosemount pressure transmitters (model 3154) OF Angra I Power Plant | pt_BR |
dc.type | Texto completo de evento | pt_BR |
dspace.entity.type | Publication | |
ipen.autor | IRACI MARTINEZ PEREIRA GONCALVES | |
ipen.autor | ROBERTO CARLOS DOS SANTOS | |
ipen.codigoautor | 1058 | |
ipen.codigoautor | 499 | |
ipen.contributor.ipenauthor | IRACI MARTINEZ PEREIRA GONCALVES | |
ipen.contributor.ipenauthor | ROBERTO CARLOS DOS SANTOS | |
ipen.date.recebimento | 18-01 | pt_BR |
ipen.event.datapadronizada | 2017 | pt_BR |
ipen.identifier.ipendoc | 24034 | pt_BR |
ipen.notas.internas | Proceedings | pt_BR |
ipen.type.genre | Artigo | |
relation.isAuthorOfPublication | 68c7c73d-2c58-4074-901c-36d1e763d624 | |
relation.isAuthorOfPublication | 327648c2-5283-4511-856f-4dcdb67d3908 | |
relation.isAuthorOfPublication.latestForDiscovery | 327648c2-5283-4511-856f-4dcdb67d3908 | |
sigepi.autor.atividade | SANTOS, ROBERTO C. DOS:499:420:S | pt_BR |
sigepi.autor.atividade | PEREIRA, IRACI M.:1058:420:N | pt_BR |