Corrosion mechanism of Ti-6Al-4V morse taper dental implants connected to 316 L stainless steel prosthetic abutment

Imagem de Miniatura
Data
2022
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Materials Today Communications
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The aim of this work was to investigate the effect of galvanic coupling between stainless steel AISI 316 L abutment type Morse taper and implant made of ASTM F1108–14 Ti-6Al-4V alloy. The assembly of the two alloys was carried out using mechanical imbrication by means of successive strikes at 0.05 J force onto the abutment inserted in the implant along the centerline. Corrosion attack at the interface of the alloys was evaluated according to the number of strikes used for joining the parts. Corrosion resistance was evaluated for the samples by open circuit potential measurements as function of time and scanning vibrating electrode technique (SVET) and scanning ion selective electrode technique (SIET) in phosphate buffer solution with pH adjusted to 3.0, and into which hydrogen peroxide was added to simulate tissue inflammatory conditions. Samples were evaluated at the cross and longitudinal sections. Results indicated that the number of strikes used in assembling affected corrosion susceptibility. The lowest amount of corrosion products was associated to the highest number of strikes used. The corrosion resistance was related to the characteristics of the crevice between the implant and the abutment.

Como referenciar
BERBEL, LARISSA O.; VIVEIROS, BARBARA V.G. de; MICELLI, ANA L.P.; NIGRO, FREDERICO; ROSSI, JESUALDO L.; COSTA, ISOLDA. Corrosion mechanism of Ti-6Al-4V morse taper dental implants connected to 316 L stainless steel prosthetic abutment. Materials Today Communications, v. 33, p. 1-13, 2022. DOI: 10.1016/j.mtcomm.2022.104583. Disponível em: http://repositorio.ipen.br/handle/123456789/33446. Acesso em: 11 May 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções