Long-term performance assessment of HPGE detectors used in the neutron activation analysis laboratory of IPEN-CNEN/SP (Brazil)

Carregando...
Imagem de Miniatura
Data
2017
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
In this work, verification data for 11 HPGe detectors from two different manufacturers and three different intrinsic configurations were analyzed in respect to the stability of both the efficiency and resolution for the 122 keV peak from 57Co and the 1332 keV peak from 60Co. The results allow a discussion about the stability of these parameters over time (in some cases, almost 15 years), their sensitivity to imminent detector failures and their performance after a failure has been corrected; moreover, the results show a clear correlation between the manufacturer or configuration and the long-term performance of the detector.

Como referenciar
ZAHN, G.S.; GENEZINI, F.A.; TICIANELLI, R.B.; SAIKI, M. Long-term performance assessment of HPGE detectors used in the neutron activation analysis laboratory of IPEN-CNEN/SP (Brazil). Applied Radiation and Isotopes, v. 125, p. 108-112, 2017. DOI: 10.1016/j.apradiso.2017.04.012. Disponível em: http://repositorio.ipen.br/handle/123456789/27591. Acesso em: 12 May 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções