Dosimetric tests of an extrapolation chamber in standard computed tomography beams

Carregando...
Imagem de Miniatura
Data
2021
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
Computed tomography (CT) diagnostic exams are responsible for the highest dose values to the patients. There-fore, the radiation doses in this procedure must be accurate. For the dosimetry of CT beams, the radiation detec-tor is usually a pencil-type ionization chamber. This type of dosimeter presents a uniform response to the inci-dent radiation beam from all angles, which makes it suitable for such equipment since the X-ray tube executes a circular movement around the table during irradiation. However, there is no primary standard system for this kind of radiation beam yet. In order to search for a CT primary standard, an extrapolation chamber built at the Calibration Laboratory (LCI) of the Instituto de Pesquisas Energéticas e Nucleares (IPEN) was tested. An ex-trapolation chamber is a parallel-plate ionization chamber that allows the variation of its sensitive air volume. This chamber was used previously for low-energy radiation beams and showed results within the international recommended limits. The aim of this work is to perform some characterization tests (saturation curve, polarity effect, ion collection efficiency and linearity of response) considering the chamber depth of 1.25 mm in the radia-tion qualities for computed tomography beams at the LCI. The results showed to be within the international recommended limits.

Como referenciar
CASTRO, M.C.; SILVA, N.F.; CALDAS, L.V.E. Dosimetric tests of an extrapolation chamber in standard computed tomography beams. Brazilian Journal of Radiation Sciences, v. 9, n. 1A, p. 1-6, 2021. DOI: 10.15392/bjrs.v9i1A.1338. Disponível em: http://repositorio.ipen.br/handle/123456789/32097. Acesso em: 11 May 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções