Dosimetric application of a special pencil ionization chamber in radiotherapy x-ray beams

Carregando...
Imagem de Miniatura
Data
2014
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radiation Physics and Chemistry
ODS
ODS 3
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The aim of this work was to study the performance of a pencil ionization chamber with a sensitive volume of only 1.06 cm3 and a length of 3.0 cm, developed at the Calibration Laboratory of the IPEN, in very low-energy radiotherapy X-ray beams. These beams are still used for certain skin cancer treatments due to their rapid attenuation in tissue. The dosimeter performance was evaluated in some tests proposed by the IEC 60731 standard: short- and long-term stability and linearity of response. For a complete analysis of the dosimeter response, the EGSnrc Monte Carlo simulation was utilized to investigate the influence of its different parts on the ionization chamber response. All results of the tests were in accordance with the recommended limits, and this work shows that it is possible to extend the application of this pencil-type ionization chamber developed at the LCI.

Como referenciar
NEVES, LUCIO P.; PERINI, ANA P.; FERNANDEZ VAREA, JOSÉ M.; CASSOLA, VAGNER F.; KRAMER, RICHARD; KHOURY, HELEN J.; CALDAS, LINDA V.E. Dosimetric application of a special pencil ionization chamber in radiotherapy x-ray beams. Radiation Physics and Chemistry, v. 95, p. 98-100, 2014. DOI: 10.1016/j.radphyschem.2012.12.042. Disponível em: http://repositorio.ipen.br/handle/123456789/8547. Acesso em: 05 Jun 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções