Determination of transmission factors in beta radiation beams

Carregando...
Imagem de Miniatura
Data
2018
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
In beta emitters, in order to evaluate the absorbed dose rate at different tissue depths, it is necessary to determine the transmission factors. In this work, the transmission factors determined in beta secondary standard radiation beams are presented. For this purpose, an extrapolation chamber was used. The results obtained were considered acceptable, and they are within the uncertainties in comparison with the values provided by the source calibration certificate. The maximum differences between the results obtained in this work and those from the calibration certificate were 3.3%, 3.8% and 5.9% for 90Sr/90Y, 85Kr and 147Pm sources respectively.

Como referenciar
POLO, IVON O.; CALDAS, LINDA V.E. Determination of transmission factors in beta radiation beams. Applied Radiation and Isotopes, v. 136, p. 82-86, 2018. DOI: 10.1016/j.apradiso.2018.02.010. Disponível em: http://repositorio.ipen.br/handle/123456789/28890. Acesso em: 08 May 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções