Comparison for air kerma from radiation protection gamma-ray beams with brazilian network - 2016/2017

Carregando...
Imagem de Miniatura
Data
2018
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Journal of Physics: Conference Series
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The results of the comparison involving 9 laboratories in Brazil are reported. The measured quantity was the air kerma in 137Cs and 60Co, at the level of radioprotection. The comparison was conducted by the National Laboratory Metrology of Ionizing Radiation (LNMRI/IRD) from October 2016 to March 2017. The largest deviation between the calibration coefficients was 0.8% for 137Cs and 0.7% for 60Co. This proficiency exercise proved the technical capacity of the Brazilian calibration network in radiation monitors and the results were used by some in the implementation of the standard ISO/IEC 17025.

Como referenciar
CABRAL, T.S.; SILVA, C.N.M. da; POTIENS, M.P.A.; SOARES, C.M.A.; SILVEIRA, R.R.; KHOURY, H.; SAITO, V.; FERNANDES, E.; CARDOSO, W.F.; OLIVEIRA, H.P.S. de; PIRES, M.A.; AMORIM, A.S. de; BALTHAR, M. Comparison for air kerma from radiation protection gamma-ray beams with brazilian network - 2016/2017. Journal of Physics: Conference Series, v. 975, p. 1-9, 2018. DOI: 10.1088/1742-6596/975/1/012042. Disponível em: http://repositorio.ipen.br/handle/123456789/29469. Acesso em: 26 Apr 2024.