Comparison for air kerma from radiation protection gamma-ray beams with brazilian network - 2016/2017
Carregando...
Data
2018
Data de publicação:
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Journal of Physics: Conference Series
Resumo
The results of the comparison involving 9 laboratories in Brazil are reported. The
measured quantity was the air kerma in 137Cs and 60Co, at the level of radioprotection. The
comparison was conducted by the National Laboratory Metrology of Ionizing Radiation
(LNMRI/IRD) from October 2016 to March 2017. The largest deviation between the
calibration coefficients was 0.8% for 137Cs and 0.7% for 60Co. This proficiency exercise proved
the technical capacity of the Brazilian calibration network in radiation monitors and the results
were used by some in the implementation of the standard ISO/IEC 17025.
Como referenciar
CABRAL, T.S.; SILVA, C.N.M. da; POTIENS, M.P.A.; SOARES, C.M.A.; SILVEIRA, R.R.; KHOURY, H.; SAITO, V.; FERNANDES, E.; CARDOSO, W.F.; OLIVEIRA, H.P.S. de; PIRES, M.A.; AMORIM, A.S. de; BALTHAR, M. Comparison for air kerma from radiation protection gamma-ray beams with brazilian network - 2016/2017. Journal of Physics: Conference Series, v. 975, p. 1-9, 2018. DOI: 10.1088/1742-6596/975/1/012042. Disponível em: http://repositorio.ipen.br/handle/123456789/29469. Acesso em: 26 Apr 2024.