Some characterization tests for an extrapolation chamber in CT standard beams in a specific chamber depth

Carregando...
Imagem de Miniatura
Data
2020
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
Among all diagnostic exams, the computed tomography (CT) is responsible for the highest dose values to the patients. So, the radiation doses in this procedure must be accurate for the dosimetry procedure in CT scanner beams making use of pencil ionization chambers with sensitive volume lengths of 10 cm. However, there is no primary standard system for this kind of radiation beam. An extrapolation ionization chamber, built at the Cali-bration Laboratory (LCI) was tested to establish a CT primary standard. The aim of this work was to perform some characterization tests (short- and medium-term stabilities, energy dependence, angular dependence) in the standard X-ray beams established for computed tomography at the LCI in a specific chamber depth (1.25 mm). The results showed to be within the international recommended limits except for the energy dependence.

Como referenciar
CASTRO, M.C.; SILVA, N.F.; CALDAS, L.V.E. Some characterization tests for an extrapolation chamber in CT standard beams in a specific chamber depth. Brazilian Journal of Radiation Sciences, v. 8, n. 1A, p. 1-6, 2020. DOI: 10.15392/bjrs.v8i1A.1085. Disponível em: http://200.136.52.105/handle/123456789/31598. Acesso em: 10 May 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções