The performance of a multi guard ring (MGR) diode for clinical electron beams dosimetry
dc.contributor.author | NASCIMENTO, C.R. | |
dc.contributor.author | ASFORA, V.K. | |
dc.contributor.author | GONCALVES, J.A.C. | |
dc.contributor.author | KHOURY, H.J. | |
dc.contributor.author | BARROS, V.S.M. | |
dc.contributor.author | KALIL, L.F. | |
dc.contributor.author | BUENO, C.C. | |
dc.coverage | Internacional | pt_BR |
dc.date.accessioned | 2018-12-06T16:16:46Z | |
dc.date.available | 2018-12-06T16:16:46Z | |
dc.date.issued | 2018 | pt_BR |
dc.description.abstract | The dosimetric response of a multi guard ring structure (MGR) diode has been studied with clinical electron beam energies from 5 MeV to 15 MeV. The results showed that the MGR dose response is linear in the range of 5–320 cGy and presents reproducibility with variation coefficients less than 0.4%. The field output factors measured with the MGR agreed within 2% with those measured with an ionization chamber. This study evidences that this diode can be used for clinical electron beam dosimetry. | pt_BR |
dc.description.sponsorship | Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) | pt_BR |
dc.description.sponsorshipID | CNPq: 305903/2011-0; 310668/2012-3 | pt_BR |
dc.format.extent | 112-117 | pt_BR |
dc.identifier.citation | NASCIMENTO, C.R.; ASFORA, V.K.; GONCALVES, J.A.C.; KHOURY, H.J.; BARROS, V.S.M.; KALIL, L.F.; BUENO, C.C. The performance of a multi guard ring (MGR) diode for clinical electron beams dosimetry. <b>Applied Radiation and Isotopes</b>, v. 141, p. 112-117, 2018. DOI: <a href="https://dx.doi.org/10.1016/j.apradiso.2018.07.002">10.1016/j.apradiso.2018.07.002</a>. Disponível em: http://repositorio.ipen.br/handle/123456789/29331. | |
dc.identifier.doi | 10.1016/j.apradiso.2018.07.002 | pt_BR |
dc.identifier.issn | 0969-8043 | pt_BR |
dc.identifier.orcid | aguardando | pt_BR |
dc.identifier.orcid | https://orcid.org/0000-0001-7881-7254 | |
dc.identifier.orcid | https://orcid.org/0000-0002-8940-9544 | |
dc.identifier.percentilfi | 38.22 | pt_BR |
dc.identifier.percentilfiCiteScore | 53.00 | |
dc.identifier.uri | http://repositorio.ipen.br/handle/123456789/29331 | |
dc.identifier.vol | 141 | pt_BR |
dc.relation.ispartof | Applied Radiation and Isotopes | pt_BR |
dc.rights | openAccess | pt_BR |
dc.subject | electron beams | |
dc.subject | electron dosimetry | |
dc.subject | ionization chambers | |
dc.subject | mev range 01-10 | |
dc.subject | mev range 10-100 | |
dc.subject | rings | |
dc.subject | security personnel | |
dc.subject | semiconductor detectors | |
dc.subject | semiconductor materials | |
dc.title | The performance of a multi guard ring (MGR) diode for clinical electron beams dosimetry | pt_BR |
dc.type | Artigo de periódico | pt_BR |
dspace.entity.type | Publication | |
ipen.autor | CARMEN CECILIA BUENO TOBIAS | |
ipen.autor | JOSEMARY ANGELICA CORREA GONCALVES | |
ipen.codigoautor | 1592 | |
ipen.codigoautor | 924 | |
ipen.contributor.ipenauthor | CARMEN CECILIA BUENO TOBIAS | |
ipen.contributor.ipenauthor | JOSEMARY ANGELICA CORREA GONCALVES | |
ipen.date.recebimento | 18-12 | pt_BR |
ipen.identifier.fi | 1.343 | pt_BR |
ipen.identifier.fiCiteScore | 2.1 | |
ipen.identifier.ipendoc | 25124 | pt_BR |
ipen.identifier.iwos | WoS | pt_BR |
ipen.range.fi | 0.001 - 1.499 | |
ipen.range.percentilfi | 25.00 - 49.99 | |
ipen.type.genre | Artigo | |
relation.isAuthorOfPublication | fa74399b-83a0-4f46-91e0-da469104d3f6 | |
relation.isAuthorOfPublication | 76fdc4d1-7624-4332-a9d0-f06826000679 | |
relation.isAuthorOfPublication.latestForDiscovery | 76fdc4d1-7624-4332-a9d0-f06826000679 | |
sigepi.autor.atividade | GONCALVES, J.A.C.:924:240:N | pt_BR |
sigepi.autor.atividade | BUENO, C.C.:1592:240:N | pt_BR |