In this paper, various simpli cations of the HDR source Varian VariSource Classic model, in which 192Ir as a radionuclide is used, were compared. These simpli cations were carried out by Monte Carlo simulations, using the MCNPX 2.7.0 code. The different sources were compared through a distribution of energy deposition in a water phantom. Our results indicated that small simpli cations will present no in uence on the source response, and the removal of the entire capsule surrounding the radionuclide will present a difference of just 0.53% in the nal response.

Como referenciar
SANTOS, W.S.; GONALVES, P.E.; BELINATO, W.; CALDAS, L.V.E.; PERINI, A.P.; NEVES, L.P. The influence of different 192Ir sources geometries to the energy deposition. Journal of Physics: Conference Series, v. 733, p. 1-5, 2016. DOI: 10.1088/1742-6596/733/1/012092. Disponível em: http://repositorio.ipen.br/handle/123456789/27805. Acesso em: 08 May 2024.