Stability study of the alanine epr dosimetry system at IPEN/CNEN SP

Carregando...
Imagem de Miniatura
Data
Data de publicação:
2019
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The dosimetry system using alanine EPR is a standard secondary calibration system for high doses. IPEN's High Doses Dosimetry Laboratory (LDA) has an Electron Paramagnetic Resonance equipment operating in the X band which performs high dose dosimetry service for electron and X ray fields for the dose range of 10Gy to 200kGy. IPEN has established a dosimetry system based on an alanine dosimeter developed at the institute. The system is composed of pure DL alanine encapsulated in a polyethylene microtube, which allows estimated doses of 10Gy to 200kGy. The advantages of this system are the low signal fading over a long period of time (over 90 days) and dose measurements above 150kGy. The main disadvantage is the difficulty in measures below 10Gy. In order to improve quality assurance services delivered in radiotherapy, commercial alanine dosimeters are being tested and compared with the already established system for measurements with time intervals longer than 90 days. The present work is an evaluation of the commercial L alanine dosimeter signal compared to the alanine dosimeter produced in IPEN. The dosimeters were irradiated with X ray and E beam and doses of 1, 10 and 20Gy. All measurements were performed following the same laboratory protocol based in ISO/ASTM guidelines. Environmental effects such as temperature, sensitivity to light and air humidity, and other storage conditions were evaluated. The results show that the decay of commercial dosimeters over 36 months was 11% higher than that observed in the dosimeter developed in IPEN, indicating the need for a correction factor for the comparison of dosimetry systems.

Como referenciar
RODRIGUES JR ., ORLANDO; SANTOS, SILAS C. dos. Stability study of the alanine epr dosimetry system at IPEN/CNEN SP. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 21-25, 2019, Santos, SP. Proceedings... Rio de Janeiro: Associação Brasileira de Energia Nuclear, 2019. p. 259-264. Disponível em: http://repositorio.ipen.br/handle/123456789/30350. Acesso em: 06 Feb 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento
Coleções