Count rate effect on the response of a low-cost PIN diode for electron spectrometry
Carregando...
Data
Data de publicação:
2019
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
Resumo
The response of a low-cost Si photodiode model BPX 65 for low-energy electron spectrometry is investigated
envisaging its use in measurements of electron multiple elastic scattering. The electron beam with energy
between 10-100 keV is delivered by the gun of the Racetrack Microtron at Instituto de Física, Universidade de
São Paulo, with an energy dispersion less than 0.5 keV. The energy resolution achieved was less than 3.5 keV,
limited mainly by noise from the electronic acquisition chain. For count rates between 20 and 5500 counts/s, the
variation on the centroid of electron peak was smaller than 0.4% throughout the energy range. Therefore, the
BPX 65 is suitable for electron spectrometry.
Como referenciar
PETRI, ANNA R.; BARROS, SUELEN F.; GONCALVES, JOSEMARY A.C.; BUENO, CARMEN C.; MAIDANA, NORA; MARTINS, MARCOS N.; VANIN, VITO R. Count rate effect on the response of a low-cost PIN diode for electron spectrometry. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 21-25, 2019, Santos, SP. Proceedings... Rio de Janeiro: Associação Brasileira de Energia Nuclear, 2019. p. 1207-1213. Disponível em: http://repositorio.ipen.br/handle/123456789/30559. Acesso em: 06 Feb 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.