New model for an epoxy-based brachytherapy source to be used in spinal cancer treatment

Carregando...
Imagem de Miniatura
Data
2021
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
ODS
ODS 3
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The present work described the cold fabrication of a P-32 radioactive source to be used in CNS cancer using epoxy resin. The epoxy plaque fabricated with Teflon mold presented better agreement. MCNP simulation evaluated the radiation dose. Special attention was given to factors that can impact dose distribution. Average dose was 16.44 ± 2.89% cGy/s. Differences of less than 0.01 cm in thickness within the plaque lead to differences of up to 12% in the dose rate.

Como referenciar
SILVA, JOSE T.; SOUZA, CARLA D. de; ANGELOCCI, LUCAS V.; ROSERO, WILMMER A.A.; NOGUEIRA, BEATRIZ R.; CORREIA, RUANYTO W.; ZEITUNI, CARLOS A.; ROSTELATO, MARIA E.C.M. New model for an epoxy-based brachytherapy source to be used in spinal cancer treatment. Applied Radiation and Isotopes, v. 178, p. 1-7, 2021. DOI: 10.1016/j.apradiso.2021.109952. Disponível em: http://repositorio.ipen.br/handle/123456789/32576. Acesso em: 10 May 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.