Verification of angular dependence in mosfet detector

Carregando...
Imagem de Miniatura

Data

Data de publicação

2017

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
In vivo dosimetry is an essential tool for quality assurance programs, being a procedure commonly performed with thermoluminescent dosimeters (TLDs) or diodes. However, a type of dosimeter that has increasing popularity in recent years is the metal-oxide-semiconductor field effect transistor (MOSFET) detector. MOSFET dosimeters fulfill all the necessary characteristics to realize in vivo dosimetry since it has a small size, good precision and feasibility of measurement, as well as easy handling. Nevertheless, its true differential is to allow reading of the dose in real time, enabling immediate intervention in the correction of physical parameters deviations and anticipation of small anatomical changes in a patient during treatment. In order for MOSFET dosimeter to be better accepted in clinical routine, information reporting performance should be available frequently. For this reason, this work proposes to verify reproducibility and angular dependence of a standard sensitivity MOSFET dosimeter (TN-502RD-H) for Cs-137 and Co-60 sources. Experimental data were satisfactory and MOSFET dosimeter presented a reproducibility of 3.3% and 2.7% (1 SD) for Cs-137 and Co-60 sources, respectively. In addition, an angular dependence of up to 6.1% and 16.3% for both radioactive sources, respectively. It is conclusive that MOSFET dosimeter TN-502RD-H has satisfactory reproducibility and a considerable angular dependence, mainly for the Co-60 source. This means that although precise measurements, special attention must be taken for applications in certain anatomical regions in a patient.

Como referenciar
SOUZA, CLAYTON H.; SHORTO, JULIAN M.B.; SIQUEIRA, PAULO T.D.; NUNES, MAIRA G.; SILVA JUNIOR, IREMAR A.; YORIYAZ, HELIO. Verification of angular dependence in mosfet detector. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 22-27, 2017, Belo Horizonte, MG. Proceedings... Rio de Janeiro, RJ: Associação Brasileira de Energia Nuclear, 2017. Disponível em: http://repositorio.ipen.br/handle/123456789/28228. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções