Machine learning in PSQA

dc.contributor.authorALVES, NATHALIA L. A.
dc.contributor.authorMENEZES, MARIO O. de
dc.contributor.authorYORIYAZ, HELIO
dc.contributor.authorRADICCHI, LUCAS A.
dc.contributor.authorMATTESCO, ANDRE L. S.
dc.contributor.authorPROSPERO, ANDRE G.
dc.coverageInternacional
dc.creator.eventoINTERNATIONAL SYMPOSIUM ON SOLID STATE DOSIMETRY, 25th
dc.date.accessioned2026-11-19T11:44:11Z
dc.date.available2026-11-19T11:44:11Z
dc.date.eventoSeptember 22-26, 2025
dc.description.abstractThis study explores the application of supervised machine learning (ML) algorithms to predict outcomes of patient-specific quality assurance (PSQA) in intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). A cohort of 133 anonymized clinical radiotherapy plans covering treatment sites such as prostate, head and neck, pelvis, central nervous system (CNS), and stereotactic radiosurgery (SRS) was analyzed. From each DICOM-RT (Digital Imaging and Communications in Medicine – Radiation Therapy) plan, 60 quantitative features were extracted, encompassing multileaf collimator (MLC) dynamics, dose distribution characteristics, and complexity indices including Modulation Complexity Score (MCS), Average Leaf Speed (ALS), and Leaf Sequence Variability (LSV). To address class imbalance between QA pass/fail outcomes, a synthetic dataset of 66 perturbed plans simulating realistic delivery deviations was generated, complemented by the use of the Synthetic Minority Oversampling Technique (SMOTE) [8]. Feature importance analysis identified MCS, ALS, and dose irregularity metrics as key predictors of QA failure. The optimized Random Forest classifier achieved a weighted F1-score of 0.99 during cross-validation and 0.92 on the independent test set. The results corroborate findings in the literature that support the integration of ML models to enhance PSQA accuracy and sensitivity beyond conventional gamma-based analysis [1, 10, 7, 9]. By leveraging DICOM-RT metadata, this study highlights the feasibility of implementing predictive algorithms in radiotherapy QA pipelines, improving workflow efficiency and reducing dependence on manual dosimetric verification.
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
dc.description.sponsorshipIDCNPq: 406761/2022-1
dc.event.siglaISSSD
dc.format.extent220-229
dc.identifier.citationALVES, NATHALIA L. A.; MENEZES, MARIO O. de; YORIYAZ, HELIO; RADICCHI, LUCAS A.; MATTESCO, ANDRE L. S.; PROSPERO, ANDRE G. Machine learning in PSQA: performance of a random forest model on radiotherapy metrics. In: INTERNATIONAL SYMPOSIUM ON SOLID STATE DOSIMETRY, 25th, September 22-26, 2025, Belo Horizonte, MG. <b>Proceedings...</b> p. 220-229. Disponível em: https://repositorio.ipen.br/handle/123456789/50293.
dc.identifier.orcidhttps://orcid.org/0000-0003-0263-3541
dc.identifier.orcidhttps://orcid.org/0000-0003-4023-236X
dc.identifier.urihttps://repositorio.ipen.br/handle/123456789/50293
dc.language.isoeng
dc.local.eventoBelo Horizonte, MG
dc.rightsopenAccess
dc.titleMachine learning in PSQA
dc.typeTexto completo de evento
dspace.entity.typePublication
ipen.autorNATHALIA LUZIA AIDAR ALVES
ipen.autorMARIO OLIMPIO DE MENEZES
ipen.autorHELIO YORIYAZ
ipen.codigoautor16251
ipen.codigoautor699
ipen.codigoautor80
ipen.contributor.ipenauthorNATHALIA LUZIA AIDAR ALVES
ipen.contributor.ipenauthorMARIO OLIMPIO DE MENEZES
ipen.contributor.ipenauthorHELIO YORIYAZ
ipen.event.datapadronizada2025
ipen.identifier.ipendoc32304
ipen.notas.internasProceedings
ipen.subtituloperformance of a random forest model on radiotherapy metrics
ipen.type.genreArtigo
relation.isAuthorOfPublication4e431c52-4991-407f-a8e8-96b14d1d1980
relation.isAuthorOfPublication447febbf-b8c8-40fa-ac57-283881bad56e
relation.isAuthorOfPublication93102dac-f89d-4743-b6ea-5158708cbb92
relation.isAuthorOfPublication.latestForDiscovery4e431c52-4991-407f-a8e8-96b14d1d1980
sigepi.autor.atividadeNATHALIA LUZIA AIDAR ALVES:16251:-1:N
sigepi.autor.atividadeMARIO OLIMPIO DE MENEZES:699:310:N
sigepi.autor.atividadeHELIO YORIYAZ:80:420:N

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
32304.pdf
Tamanho:
218.27 KB
Formato:
Adobe Portable Document Format

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções