Application of the correction factor for radiation quality Kq in dosimetry with pencil-type ionization chambers using a tandem system

dc.contributor.authorFONTES, LADYJANE P.
dc.contributor.authorPOTIENS, MARIA da P.A.
dc.coverageInternacionalpt_BR
dc.creator.eventoINTERNATIONAL NUCLEAR ATLANTIC CONFERENCEpt_BR
dc.date.accessioned2018-01-08T10:37:00Z
dc.date.available2018-01-08T10:37:00Z
dc.date.eventoOctober 22-27, 2017pt_BR
dc.description.abstractThe pencil-type ionization chamber widely used in computed tomography (CT) dosimetry, is a measuring instrument that has a cylindrical shape and provides uniform response independent of the angle of incidence of ionizing radiation. Calibration and measurements performed with the pencil-type ionization chamber are done in terms of Kerma product in air-length (Pk,l) and values are given in Gy.cm. To obtain the values of (Pk,l) during clinical measurements, the readings performed with the ionization chamber are multiplied by the calibration coefficient (Nk,l) and the correction factor C for quality (Kq) which are given in Calibration certificates of the chambers. The application of the correction factor for radiation quality Kq is done as a function of the effective energy of the beam that is determined by the Half Value layer (HVL) calculation. In order to estimate the HVL values in this work, a Tandem system made up of cylindrical aluminum and PMMA absorber layers was used as a low cost and easy to apply method. From the Tandem curve, it was possible to construct the calibration curve and obtain the appropriate Kq to the beam of the computed tomography equipment studied.pt_BR
dc.event.siglaINACpt_BR
dc.identifier.citationFONTES, LADYJANE P.; POTIENS, MARIA da P.A. Application of the correction factor for radiation quality Kq in dosimetry with pencil-type ionization chambers using a tandem system. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 22-27, 2017, Belo Horizonte, MG. <b>Proceedings...</b> Rio de Janeiro, RJ: Associação Brasileira de Energia Nuclear, 2017. Disponível em: http://repositorio.ipen.br/handle/123456789/28232.
dc.identifier.orcidhttps://orcid.org/0000-0002-4049-6720
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/28232
dc.localRio de Janeiro, RJpt_BR
dc.local.eventoBelo Horizonte, MGpt_BR
dc.publisherAssociação Brasileira de Energia Nuclearpt_BR
dc.rightsopenAccesspt_BR
dc.subjectcalibration
dc.subjectcomputerized tomography
dc.subjectcorrections
dc.subjectdosimetry
dc.subjectionization chambers
dc.subjectkerma
dc.subjectlayers
dc.subjectpmma
dc.subjectquality assurance
dc.titleApplication of the correction factor for radiation quality Kq in dosimetry with pencil-type ionization chambers using a tandem systempt_BR
dc.typeTexto completo de eventopt_BR
dspace.entity.typePublication
ipen.autorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.autorLADYJANE PEREIRA FONTES
ipen.codigoautor346
ipen.codigoautor12137
ipen.contributor.ipenauthorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.contributor.ipenauthorLADYJANE PEREIRA FONTES
ipen.date.recebimento18-01pt_BR
ipen.event.datapadronizada2017pt_BR
ipen.identifier.ipendoc24057pt_BR
ipen.notas.internasProceedingspt_BR
ipen.type.genreArtigo
relation.isAuthorOfPublication30f213ae-321d-4d51-8763-0a2585255fb9
relation.isAuthorOfPublication00349d21-f362-4498-a5c0-691c31365fa4
relation.isAuthorOfPublication.latestForDiscovery00349d21-f362-4498-a5c0-691c31365fa4
sigepi.autor.atividadeFONTES, LADYJANE P.:12137:330:Spt_BR
sigepi.autor.atividadePOTIENS, MARIA DA P.A.:346:330:Npt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
24057.pdf
Tamanho:
418.57 KB
Formato:
Adobe Portable Document Format
Descrição:

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções