Evaluation of ionization chambers in X-ray beams used for instruments calibration

dc.contributor.authorCORREA, EDUARDO de L.pt_BR
dc.contributor.authorPOTIENS, MARIA da P.A.pt_BR
dc.contributor.authorCALDAS, LINDA V.E.pt_BR
dc.contributor.authorVIVOLO, VITORpt_BR
dc.coverageInternacionalpt_BR
dc.creator.eventoINTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATIONpt_BR
dc.date.accessioned2014-11-17T18:02:46Zpt_BR
dc.date.accessioned2014-11-18T18:15:10Zpt_BR
dc.date.accessioned2015-04-02T04:44:46Z
dc.date.available2014-11-17T18:02:46Zpt_BR
dc.date.available2014-11-18T18:15:10Zpt_BR
dc.date.available2015-04-02T04:44:46Z
dc.date.eventoOctober 19-24, 2008pt_BR
dc.description.abstractThe periodically evaluation of ionization chambers and quality control of standard X-ray beams used for purpose to calibration is important to guarantee the calibration procedures of dosimeters and instruments used in radiation protection, diagnostic radiology and radiotherapy routines to ensure that procedures are following the national and international recommendations. Based on that, a quality control routine must be made periodically, to guarantee the good working of these reference ionization chambers and X-ray beams. In this work was compared the performance of different kind of ionization chambers in X-ray beams, level mammography. It has been used as reference a 6 cm³ ionization chamber, from RadcalTM, with MylarTM window, specific for use in mammography beams. The performance of that ionization chamber was compared with other two 180 cm³ (parallel plate chambers) and two 6 cm³ (cylindrical type). The ionization chambers’ stability has been made using a 90Sr+90Y radioactive source (20 MBq, 2007). After the stability control, the ionization chambers were placed on X-ray beams, mammography level (25, 27.5 and 35 kV, and current of 30 mA), tungsten target with additional filtration of 0.06 mm of molybdenum for direct beams, and plus 2 mm of aluminum for attenuated beams. The results show that the 6 cm³ cylindrical ionization chambers have a maximum variation of 5.08% in the energies range studied, the results show that they could be used as working standard in the calibration laboratory. The 180 cm³ ionization chambers have an average variation of about 20% to direct beams, and about 7,8% to attenuated beams. It is possible conclude, that the 180 cm³ ionization chambers are not very efficient to low energies measurements, although they also can be used in diagnostic radiology X-ray qualities, but to higher energies, and can also be used in radiation protection X-ray qualities.
dc.event.siglaIRPA12pt_BR
dc.identifier.citationCORREA, EDUARDO de L.; POTIENS, MARIA da P.A.; CALDAS, LINDA V.E.; VIVOLO, VITOR. Evaluation of ionization chambers in X-ray beams used for instruments calibration. In: INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, October 19-24, 2008, Buenos Aires, Argentina. <b>Proceedings...</b> Disponível em: http://repositorio.ipen.br/handle/123456789/15940.
dc.identifier.orcidhttps://orcid.org/0000-0002-7362-2455
dc.identifier.orcidhttps://orcid.org/0000-0002-4049-6720
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/15940pt_BR
dc.local.eventoBuenos Aires, Argentinapt_BR
dc.rightsopenAccesspt_BR
dc.subjectionization chamberspt_BR
dc.subjectx-ray sourcespt_BR
dc.subjectbeamspt_BR
dc.subjectquality controlpt_BR
dc.subjectcalibrationpt_BR
dc.subjectbiomedical radiographypt_BR
dc.subjectmammary glandspt_BR
dc.subjectradiologypt_BR
dc.subjectdiagnostic techniquespt_BR
dc.titleEvaluation of ionization chambers in X-ray beams used for instruments calibrationpt_BR
dc.typeTexto completo de eventopt_BR
dspace.entity.typePublication
ipen.autorVITOR VIVOLO
ipen.autorLINDA V. E. CALDAS
ipen.autorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.autorEDUARDO DE LIMA CORREA
ipen.codigoautor406
ipen.codigoautor1495
ipen.codigoautor346
ipen.codigoautor7498
ipen.contributor.ipenauthorVITOR VIVOLO
ipen.contributor.ipenauthorLINDA V. E. CALDAS
ipen.contributor.ipenauthorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.contributor.ipenauthorEDUARDO DE LIMA CORREA
ipen.date.recebimento09-03pt_BR
ipen.event.datapadronizada2008pt_BR
ipen.identifier.ipendoc13391pt_BR
ipen.notas.internasProceedingspt_BR
ipen.type.genreArtigo
relation.isAuthorOfPublication6875184b-48f2-43e7-b6de-e676ee5db15a
relation.isAuthorOfPublication7f46d4f4-dfd6-4485-a767-10df5b4f4f13
relation.isAuthorOfPublication30f213ae-321d-4d51-8763-0a2585255fb9
relation.isAuthorOfPublication1f00087b-0c1b-413b-8464-2842b24a7328
relation.isAuthorOfPublication.latestForDiscovery1f00087b-0c1b-413b-8464-2842b24a7328
sigepi.autor.atividadeCORREA, EDUARDO de L.:7498:330:Spt_BR
sigepi.autor.atividadePOTIENS, MARIA da P.A.:346:330:Npt_BR
sigepi.autor.atividadeCALDAS, LINDA V.E.:1495:330:Npt_BR
sigepi.autor.atividadeVIVOLO, VITOR:406:330:Npt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
13391.pdf
Tamanho:
142.13 KB
Formato:
Adobe Portable Document Format

Coleções