Phantom positioning variation in the gamma knife® perfexion dosimetry

Carregando...
Imagem de Miniatura
Data
Data de publicação:
2016
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
BRAZILIAN CONGRESS ON METROLOGY, 8th
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The use of small volume ionization chamber has become required for the dosimetry of equipments that use small radiation fields such as the Gamma Knife® Perfexion (GKP) unit. In this work, a pinpoint ionization chamber was inserted into the dosimetry phantom and measurements were performed with the phantom in different positions, in order to verify if the change in the phantom positioning affects the dosimetry of the GKP. Four different phantom positions were performed. The variation in the result is within the range allowed for the dosimetry of a GKP equipment.

Como referenciar
COSTA, N.A.; POTIENS, M.P.A.; SARAIVA, C.W.C. Phantom positioning variation in the gamma knife® perfexion dosimetry. In: BRAZILIAN CONGRESS ON METROLOGY, 8th, 29 de novembro - 04 de dezembro, 2015, Bento Gonçalves, RS. Proceedings... p. 1-5. Disponível em: http://repositorio.ipen.br/handle/123456789/27223. Acesso em: 19 Feb 2025.