  • Artigo IPEN-doc 27771
    Monte Carlo simulation to assess free space and end-weld thickness variation effects on dose rate for a new Ir-192 brachytherapy source
    2021 - ANGELOCCI, LUCAS V.; SOUZA, CARLA D. de; PANTELIS, EVAGGELOS; NOGUEIRA, BEATRIZ R.; ZEITUNI, CARLOS A.; ROSTELATO, MARIA E.C.M.
    A new Iridium-192 seed for brachytherapy is under development. Specific dose rate contribution by two different factors were evaluated: the effect from movement of the core in the free space within the seed and the effect of the end-weld thickness variation. Both were investigated through use of the Monte Carlo radiation transport code MCNP6 and an in-house routine programmed with MATLAB. Differences greater than 15% compared to results from the nominal seed were found near the source, indicating a significant dose variation.