Performance tests of a special ionization chamber for X-rays in mammography energy range

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Brazilian Journal of Radiation Sciences
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
A special mammography homemade ionization chamber was developed to be applied for mammography energy range dosimetry. This chamber has a total sensitive volume of 6 cm³ and is made of a PMMA body and graphite coated collecting electrode. Performance tests as saturation, ion collection efficiency, linearity of chamber response versus air kerma rate and energy dependence were determined. The results obtained with this special homemade ionization chamber are within the limits stated in international recommendations. This chamber can be used in quality control programs of mammography energy range. All measurements were carried out at the Calibration Laboratory of IPEN.

Como referenciar
SILVA, J.O.; CALDAS, L.V.E. Performance tests of a special ionization chamber for X-rays in mammography energy range. Brazilian Journal of Radiation Sciences, v. 6, n. 2A, p. 1-14, 2018. DOI: 10.15392/bjrs.v6i2%20-%20A.501. Disponível em: http://repositorio.ipen.br/handle/123456789/30072. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções