Application of a wrist dosimeter prototype for radiation monitoring (153Sm) during a therapeutic procedure simulation

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Journal of Radiological Protection
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
Gamma and beta radiation emitting radiopharmaceuticals are handled in nuclear medicine services, and in many cases there is only individual monitoring of gamma radiation. In this paper, the results obtained using a wrist dosimeter prototype (CaSO4:Dy + Teflon pellets) show that the doses for workers occupationally exposed to beta radiation from 153Sm are not negligible. It is important that this dose is evaluated, and it has to be taken into consideration in the individual monitoring system.

Como referenciar
CECATTI, S.G.P.; GUIMARAES, M.I.C.C.; CALDAS, L.V.E. Application of a wrist dosimeter prototype for radiation monitoring (153Sm) during a therapeutic procedure simulation. Journal of Radiological Protection, v. 29, n. 4, p. 519-525, 2009. DOI: 10.1088/0952-4746/29/4/006. Disponível em: http://repositorio.ipen.br/handle/123456789/4823. Acesso em: 30 Dec 2025.