Precision and accuracy of the methodology for the determination of Pb-210 by LSC and Po-210 by alpha spectrometry using a Sr-Spec resin
Carregando...
Data
Data de publicação
2018
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL SYMPOSIUM ON NUCLEAR AND ENVIRONMENTAL RADIOCHEMICAL ANALYSIS, 13th
Resumo
Pb-210 and Po-210 were important from the radiation protection and environmental protection point
of view, due to its high toxicity. It is part of the U-238 natural series, formed by decay of the Rn-222
gas; Pb-210 has a half-live of 22.3 years and emission beta particles with 16.5 keV (80.2%) and 63.0
keV (19.8%); and Po-210 has a half-live of 138.4 days and emission alpha particle with 5.3 MeV.
The methodology for the Pb-210 and Po-210 determination is based on the purification and
concentration of the radionuclides using a Sr-Spec resin from EICHRON and performed by Liquid
Scintillation Counting (LSC) and Alpha Spectrometry, respectively. The sample solution was loaded
onto the Sr resin column preconditioned in advance with 100 mL of 2M HCl. Lead and polonium
were retained , Po-210 was stripped with 60 mL of 6M HNO3 and Pb was eluted with 60 mL of 6M
HCl. Po was spontaneously deposited on a silver disc and counted on an alpha spectrometer for
80,000 seconds. The Pb solution was evaporated three times with 5 mL portions of 65% HNO3 and
the final residue was dissolved in 10 mL of 1M HNO3. Pb was precipitated as Pb-oxalate, which was
dried and weighed to calculate the chemical recovery using the gravimetric method. The precipitate
was quantitatively transferred into a liquid scintillation vial and 1 mL of 1M HNO3 was added to
dissolve the precipitate. The solution was mixed with 15 mL of Hisafe III scintillation solution. The
final solution was counted on a 1220 Quantulus™ Ultra Low Level Liquid Scintillation Spectrometer
for 24,000 seconds.
In this paper, evaluates the accuracy and precision of the methodology of Pb-210 by LSC and Po-210
by alpha spectrometry using a separation technique with resin Sr-Spec was performed.
Como referenciar
SAUEIA, C.H.R.; NISTI, M.B.; MAZZILLI, B.P. Precision and accuracy of the methodology for the determination of Pb-210 by LSC and Po-210 by alpha spectrometry using a Sr-Spec resin. In: INTERNATIONAL SYMPOSIUM ON NUCLEAR AND ENVIRONMENTAL RADIOCHEMICAL ANALYSIS, 13th, September 17-20, 2018, Cambridge, UK. Abstract... p. 65-65. Disponível em: http://repositorio.ipen.br/handle/123456789/30101. Acesso em: 30 Dec 2025.