Constancy check of beam quality in conventional diagnostic X-ray equipment

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Applied Radiation and Isotopes
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
A tandem ionization chamber was developed for quality control programs of X-ray equipment used in conventional radiography and mammography. A methodology for the use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities was established. The application at a medical X-ray imaging facility of this established methodology is presented. The use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities is a useful method to control the performance of the X-ray equipment.

Como referenciar
COSTA, ALESSANDRO M.; BADIN, ROMULO S.; LEITE, MARINA S.; CALDAS, LINDA V.E. Constancy check of beam quality in conventional diagnostic X-ray equipment. Applied Radiation and Isotopes, v. 66, p. 1403-1405, 2008. DOI: 10.1016/j.apradiso.2008.03.007. Disponível em: http://repositorio.ipen.br/handle/123456789/7934. Acesso em: 20 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções