Evaluation of the thermally and optically stimulated response of as Italian Silicate irradiated in 60Co beams
Carregando...
Data
Data de publicação
2017
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL TOPICAL MEETING ON INDUSTRIAL RADIATION AND RADIOISOTOPE MEASUREMENT APPLICATIONS, 10th
Resumo
In this work, pellets were manufactured in a proportion of 2:1 of powdered samples of
Obsidian black:Teflon (dimensions of 6.0 mm in diameter and 0.8 mm in thickness). The
pellets were irradiated using a Gamma-Cell 220 System, model 200, Atomic Energy of
Canada LTD (60Co). The TL and OSL responses were analyzed using the reader
system composed by the TL/OSL meter Risø, model TL/OSL-DA-20, and the TSEE
response was obtained using a homemade reader system developed at the Calibration
Laboratory, at IPEN. After the TL, OSL and TSEE measurements, the pellets were
thermally treated at 400 °C during 1 h, for reutilization. The physical and chemical
characterization of powdered Obsidian was also investigated using the X-ray diffraction,
XRD (using a diffractometer Equinox 1000, Inel), scanning electron microscope, SEM, and energy dispersive X-ray spectroscopy, EDX (the last two using a scanning electron
microscope with an energy dispersive X-ray microanalyser Vega 3 SEM, Tescon)
techniques. The results proved, basically, that Obsidian is a natural glass and
composed mainly of silica (82.4%). The TL glow curve revealed a dosimetric peak at the
temperature of 220 °C, and the TSEE emission curve showed a peak at about 300 °C.
The results obtained in the reproducibility of response test were: 2.9% (TL), 3.0%
(TSEE) and 3.1% (OSL). The lower detection limits were: 48.1 Gy (TL), 18.1 Gy (TSEE)
and 79.3 Gy (OSL). The dose-response curves showed, in the case of TL, a
supralinearity behavior between only 500 Gy and 2 kGy with a following saturation of
the response. For the TSEE response, a sublinearity was seen with a following
saturation of the response. For the OSL technique, linearity could be observed in the
interval from 500 Gy to 2 kGy and then a tendency to saturation.
Como referenciar
ANTONIO, PATRICIA L.; OLIVEIRA, RAQUEL A.P.; KHOURY, HELEN J.; CALDAS, LINDA V.E. Evaluation of the thermally and optically stimulated response of as Italian Silicate irradiated in 60Co beams. In: INTERNATIONAL TOPICAL MEETING ON INDUSTRIAL RADIATION AND RADIOISOTOPE MEASUREMENT APPLICATIONS, 10th, July 09-13, 2017, Chicago, IL, USA. Abstract... p. 58-59. Disponível em: http://repositorio.ipen.br/handle/123456789/27946. Acesso em: 20 Mar 2026.