Assessment of Sr-90 in water samples
Carregando...
Data
Data de publicação
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Resumo
The aims of this study are to evaluate the precision and accuracy of three methodologies for the determination of 90Sr in water samples: Cerenkov, LSC direct method and with radiochemical separation. The performance of the methodologies was evaluated by using two scintillation counters (Quantulus and Hidex). The parameters Minimum Detectable Activity (MDA) and Figure Of Merit (FOM) were determined for each method, the precision and accuracy were checked using 90Sr standard solutions.
Como referenciar
NISTI, M.B.; SAUEIA, C.H.R.; CASTILHO, B.; MAZZILLI, B.P. Assessment of Sr-90 in water samples: precision and accuracy. Brazilian Journal of Radiation Sciences, v. 7, n. 2A, p. 1-12, 2019. DOI: 10.15392/bjrs.v7i2A.559. Disponível em: http://repositorio.ipen.br/handle/123456789/29955. Acesso em: 20 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.