Neutron flux and gama dose profile measurements and calculations along a phantom in the BNCT facility of IEA-R1 reactor
Carregando...
Data
Data de publicação
2010
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL CONGRESS ON NEUTRON CAPTURE THERAPY, 14.
Resumo
A BNCT (Boron Neutron Capture Therapy) facility has been built at IEA-R1 reactor. The on going BNCT
experiments demand the maximization of the thermal neutron component of the irradiation fi eld and the
minimization of its epithermal and fast neutron components together with the reduction of the gamma
contamination. This work was developed with the objective of evaluating whether the present radiation fi eld in
the facility is suitable for carrying on BNCT studies and to establish a work methodology with phantoms in this
facility. In order to achieve these objectives, thermal and epithermal neutron fl ux measurements, in the sample
irradiation position were performed using hyper-pure Gold activation detectors and applying the Cadmium ratio.
Absorbed dose due to gamma radiation was determined by TLD 400. A cylindrical phantom composed by acrylic
discs was designed, built and tested in the facility. DOT 3.5 computational code was used to retrieve neutron
fl uxes and the gamma dose estimates along the phantom. Although some improvements still need to be made
in the methodology under implementation, computer simulations carried out with DOT code presented a good
agreement with experimental data for most part of the evaluated profi le. In the position corresponding to about
half the length of the phantom, the following experimental values were obtained: thermal neutron fl ux (2.52 ±
0.06)·108
n.cm-2.s-1, epithermal neutron fl ux (6.2 ± 0.3)·106 n.cm-2.s-1, absorbed dose due to thermal neutrons (4.2
± 1.8) Gy and (10.1 ± 1.3) Gy due to gamma radiation. The obtained values show that the fl uxes of thermal and
epithermal neutrons fl ux are appropriate for studies in BNCT, however, dose due to gamma radiation is high,
indicating that the facility should be improved.
Como referenciar
MUNIZ, R.O.R.; COELHO, P.R.P.; SILVA, G.S.A.; SIQUEIRA, P.T.D.; SOUSA, G.S. Neutron flux and gama dose profile measurements and calculations along a phantom in the BNCT facility of IEA-R1 reactor. In: INTERNATIONAL CONGRESS ON NEUTRON CAPTURE THERAPY, 14., Octobre 25-29, 2010, Buenos Aires, Argentina. Proceedings... Disponível em: http://repositorio.ipen.br/handle/123456789/12553. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.