Characterization of a portable thermal neutron detector
Carregando...
Data
Data de publicação
2019
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
Resumo
A portable thermal neutron detector prototype, using a PIN-type silicon photodiode coupled to a boron converter, was developed at the Nuclear and Energy Research Institute (IPEN-CNEN/SP). Several boron layers were made by Pulsed Laser Deposition method and two different prototypes were made using distinct approaches in the associated electronics. The prototypes were characterized by measurements with the cold neutron beam at PHADES (Polarized 3He and Detector Experiment Station), in the NIST Center for Neutron Research. The distance between the prototype and the neutron source was different for each prototype: 70 cm (prototype one) and 7 cm (prototype two). The linearity behavior was verified for both prototypes in order to verify the relationship between counts and neutron fluence. The intrinsic efficiency values obtained for prototypes one and two were, respectively, (1.78 ± 0.01)% and (5.2 ± 0.4)%. The angular dependence was verified only for prototype one. The concept of this detector can be applied in a future dosimeter project.
Como referenciar
COSTA, PRISCILA; RAELE, MARCUS P.; DOMIENIKAN, CLAUDIO; MADI FILHO, TUFIC; ZAHN, GUILHERME S.; GENEZINI, FREDERICO A. Characterization of a portable thermal neutron detector. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 21-25, 2019, Santos, SP. Proceedings... Rio de Janeiro: Associação Brasileira de Energia Nuclear, 2019. p. 1163-1170. Disponível em: http://repositorio.ipen.br/handle/123456789/30554. Acesso em: 20 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.