Performance of a pencil ionization chamber in various radiation beams

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Applied Radiation and Isotopes
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
Pencil ionization chambers were recommended for use exclusively in the computed tomography (CT) dosimetry, and, from the start, they were developed only with this application in view. In this work, we studied the behavior of a pencil ionization chamber in various radiation beams with the objective of extending its application. Stability tests were performed, and calibration coefficients were obtained for several standard radiation qualities of the therapeutical and diagnostic levels. The results show that the pencil ionization chamber can be used in several radiation beams other than those used in CT.

Como referenciar
MAIA, A.F.; CALDAS, L.V.E. Performance of a pencil ionization chamber in various radiation beams. Applied Radiation and Isotopes, v. 58, n. 5, p. 595-601, 2003. DOI: 10.1016/S0969-8043(03)00058-7. Disponível em: http://repositorio.ipen.br/handle/123456789/5886. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções