Performance of a pencil ionization chamber in various radiation beams
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
Resumo
Pencil ionization chambers were recommended for use exclusively in the computed tomography (CT) dosimetry, and, from the start, they were developed only with this application in view. In this work, we studied the behavior of a pencil ionization chamber in various radiation beams with the objective of extending its application. Stability tests were performed, and calibration coefficients were obtained for several standard radiation qualities of the therapeutical and diagnostic levels. The results show that the pencil ionization chamber can be used in several radiation beams other than those used in CT.
Como referenciar
MAIA, A.F.; CALDAS, L.V.E. Performance of a pencil ionization chamber in various radiation beams. Applied Radiation and Isotopes, v. 58, n. 5, p. 595-601, 2003. DOI: 10.1016/S0969-8043(03)00058-7. Disponível em: http://repositorio.ipen.br/handle/123456789/5886. Acesso em: 30 Dec 2025.