Resumo
The procedure for the standardization of 59Fe using a 4π(PC)β–γ software coincidence system is de-scribed. The standardization was performed with an experimental setup consisting of a thin window gas- flow proportional counter (PC)in4π geometry coupled to a NaI(Tl) scintillator and to a HPGe detector. The data acquisition was carried out by means of a Software Coincidence System (SCS).The beta efficiency was changed by using Collodion films Colldion films andaluminum foils as external absorbers.

Como referenciar
KOSKINAS, M.F.; POLILLO, G.; BRANCACCIO, F.; YAMAZAKI, I.M.; DIAS, M.S. Standardization of 59Fe by 4π(PC)β–γ software coincidence system. Applied Radiation and Isotopes, v. 109, p. 386-388, 2016. DOI: 10.1016/j.apradiso.2015.12.023. Disponível em: http://repositorio.ipen.br/handle/123456789/27172. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.