Utilization of Tandem systems for the detection of impurities in radiopharmaceuticals

Carregando...
Imagem de Miniatura

Data

Data de publicação

2004

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, 11th
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
In this work tandem systems were established and characterized using four cylindrical absorbers of materials with different densities for an additional quality control test. The study was performed using two secondary standard systems NPL and a homemade system composed by a well-type ionization chamber made at IPEN and a Keithley electrometer. For the establishment of the tandem systems non sealed sources were utilized. The samples with impurities were measured using combinations with different proportions of radionuclides. The ratio between the measurements with different absorbers presented constant values for each dose calibrator and pure source. The results obtained with the materials with impurities using the absorbers presented constant, but different values, depending on the impurity and its proportion in the mixture. Therefore, the use of tandem systems may be recommended for quality control programs.

Como referenciar
CECATTI, S.G.P.; CALDAS, L.V.E. Utilization of Tandem systems for the detection of impurities in radiopharmaceuticals. In: INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, 11th, 23-28 May, 2004, Madrid, Spain. Disponível em: http://repositorio.ipen.br/handle/123456789/17521. Acesso em: 26 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções