Dose calibrator and gamma counter

Carregando...
Imagem de Miniatura

Data

Data de publicação

2019

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
To ensure that pharmaceutical products have and maintain the structure, identity, purity, concentration, potency and safety characteristics required for their use, there is a set of standard procedures called Good Manufacturing Practices (GMP). In Brazil, the National Sanitary Surveillance Agency (ANVISA) regulates the GMP production of medicines through RDC17/2010 and for radiopharmaceuticals through RDC 63/2009 and 38/2008, to ensure their safe and correct use in commercial production and nuclear medicine services, respectively. Most 99mTc radiopharmaceutical monographs have biological distribution specifications. Using invasive method, 99mTcradiopharmaceuticalsare assessed by the injection into animals of defined strains and the radioactivity (as percentage of retainedor injected dose -%RDor%ID)ismeasured in specified organs. Technetium-99m Sestamibi radiopharmaceutical (99mTc-2-methoxy-isobutyl-isonitrile; 99mTc-sestamibi; MIBI-TEC®) monograph is related in USP 41, but the biological distribution assay is not included. At IPEN, the biodistribution test is performed as established in the Radiopharmaceutical Quality Control Manual of ARCAL XV (1999) - International Atomic Energy Agency (IAEA), for each batch. The objective of this work is to compare MIBI-TEC®biodistribution results using dose calibrator and gamma counter with sodium iodide detector by measuring the radioactivity in the organs. Three batches of MIBI-TEC®were used and 1 vial of lyophilized reagent (LR) was labeled with 5-10 mCiin 1-3 mL of 99mTc eluate. 200-300 μCi in 0.2 mLwereinjected in 3 Balb-Cmice. After 30 minutes of biodistribution, heart, lung, muscle, liver, paw, column and tail were withdrawn, weighed and radioactivity was measured in dose calibrator and gamma counter (μCi andcpm, respectively). %ID/g ratio in the organs of interest was calculated using the data obtained by both equipment and the results were compared. No significant differences were observed and it was possible to conclude that either a dose calibrator or a gamma counter can be used in the routine of quality control.

Como referenciar
SILVA, VITOR M.A. da; SOUSA, JURANDIR A. de; MARTINS, PATRICIA de A.; FUKUMORI, NEUZA T.O.; SILVA, NATANAEL G. da; MATSUDA, MARGARETH M.N. Dose calibrator and gamma counter: comparison of results in MIBI-TEC® biological distribution. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 21-25, 2019, Santos, SP. Proceedings... Rio de Janeiro: Associação Brasileira de Energia Nuclear, 2019. p. 1860-1864. Disponível em: http://repositorio.ipen.br/handle/123456789/30589. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções