Traceability from governamental producers of radiopharmaceuticals in measuring 18F in Brazil
Carregando...
Data
Data de publicação
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
Resumo
Since the inception of its proficiency test program to evaluate radionuclide measurement in hospitals and clinics, the National Metrology Laboratory of Ionizing Radiation-LNMRI, that represents Brazilian National Metrology Institute (NMI) for ionizing radiation has expanded its measurement and calibration capability. Requirements from the National Health Surveillance Agency from Ministry of Health (ANVISA), to producers of radiopharmaceuticals provided an opportunity to improve the full traceability chain to the highest level. Fluorodeoxyglucose (FDG-18F) is the only radiopharmaceutical simultaneously produced by all Brazilian radiopharmaceutical production centers (RPCs). By running this proficiency test, LNMRI began to provide them with the required traceability. For evaluation, the ratio of RPC to reference value results and ISO/IEC17043:2010 criteria were used. The reference value established as calibration factor on the secondary standard ionization chamber was obtained from three absolute measurements systems, and routinely confirmed in each round of proficiency test by CIEMAT/NIST liquid scintillation counting. The γ-emitting impurities were checked using a High-Purity Germanium (HPGe) detector. The results show that Brazilian RPCs are in accordance with (accuracy within 710%) the Brazilian standard for evaluation of measurements with radionuclide calibrators (CNEN NN 3.05., 2013). Nevertheless, the RPCs should improve the methodology of uncertainty estimates, essential when using the statistical criteria of ISO/IEC 17043 standard, in addition to improving accuracy to levels consistent with their position in the national traceability chain
Como referenciar
OLIVEIRA, A.E.; IWAHARA, A.; SILVA, C.J.; CRUZ, P.A.L.; POLEDNA, R.; SILVA, R.L.; LARANJEIRA, A.S.; DELGADO, J.U.; TAUHATA, L.; LOUREIRO, J.S.; TOLEDO, B.C.; BRAGHIROLLI, A.M.S.; ANDRADE, E.A.L.; SILVA, J.L.; HERNANDES, H.O.K.; VALENTE, E.S.; DALLE, H.M.; ALMEIDA, V.M.; SILVA, T.G.; FRAGOSO, M.C.F.; OLIVEIRA, M.L.; NASCIMENTO, E.S.S.; OLIVEIRA, E.M.; HERRERIAS, R.; SOUZA, A.A.; BAMBALAS, E.; BRUZINGA, W.A. Traceability from governamental producers of radiopharmaceuticals in measuring 18F in Brazil. Applied Radiation and Isotopes, v. 109, p. 236-241, 2016. DOI: 10.1016/j.apradiso.2015.11.051. Disponível em: http://repositorio.ipen.br/handle/123456789/27188. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.