Polycarbonate chromatography column to be used in a 99Mo/99mTc generator irradiated in saline solution with EB and gamma-Rays

Carregando...
Imagem de Miniatura

Data

Data de publicação

2017

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL CONFERENCE ON APPLICATIONS OF RADIATION SCIENCE AND TECHNOLOGY
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The technetium-99m generator (technetium cow or moly cow) is a device used to extract the metastable technetium isotope 99mTc from a source of decaying molybdenum-99. 99Mo has a half-life of 66 hours and can be easily transported over long distances to hospitals whereas its 99mTc decay product (half-life of 6 h is inconvenient for transport) is extracted and used for several nuclear medicine diagnostic procedures, where its short half-life is well adapted. This device works as a closed system, where the principal component is a chromatographic column of acid alumina (Al2O3) as stationary phase. Currently this column is produced using borosilicate type 1 glass. The goal of this study was to characterize the irradiated polycarbonate (PC) column in saline solution to simulate a 99Mo/99mTc generator in real use conditions. PC column and PC sampleswere EB and -irradiated in saline solution with radiation absorbed doses up to 200 kGy. Samples were analyzed by electron paramagnetic resonance spectroscopy (EPR), infrared spectroscopy (FTIR), ultraviolet spectroscopy (UV), differential scanning calorimetry (DSC) and wide-angle X-ray diffraction (WAXD). Additionally, the -irradiated PC column in presence of saline solution was studied using high performance liquid chromatography (HPLC) coupled with fluorescence detection in order to investigate the chemical phase diffusion of bisphenol A (BPA). EPR results showed at X band region a strong singlet attributed to a phenoxyl radical in the irradiated sample at room temperature. Decay of radical occurs approximately within 40 days. UV spectra presented increase in relative absorbance at 400–450 nm with increasing radiation dose. Ionizing radiation caused greenness of the original clear PC samples. This discolouration confirms the formation of phenoxyl radicals. On the other hand, glass transition temperature decreased by 1% (5K) for the maximum radiation applied dose. Similarly, a small decrease on carboxyl group peak at 1770 cm 1 was observed by FTIR. No detectable change on crystallinity was observed by WAXD. For sterilization absorbed dose, results shown no significant changes on the studied properties that way can be recommend to use PC columns instead borosilicate glass column in the 99mTc generator. The protocol via HPLC with fluorescence detection used in this work can be employed to detect the chemical phase diffusion of BPA in saline solution at ppb concentration

Como referenciar
KODAMA, Y.; CARNEIRO, R.; ROSSI, M.; RODRIGUES JUNIOR, O.; VASQUEZ, P. Polycarbonate chromatography column to be used in a 99Mo/99mTc generator irradiated in saline solution with EB and gamma-Rays. In: INTERNATIONAL CONFERENCE ON APPLICATIONS OF RADIATION SCIENCE AND TECHNOLOGY, April 24-28, 2017, Vienna, Austria. Abstract... Vienna: IAEA, 2017. p. 99-99. Disponível em: http://repositorio.ipen.br/handle/123456789/27766. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento