Dosimetric characterization of 3D printed for 137Cs gamma rays

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Journal of Physics: Conference Series
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The aim this paper is characterize materials for 3D printed with different infill percentages for 137Cs gamma rays. The RAISE 3D PRO2 printer was used to print PLA and ABS plates. Using a 137Cs source, the attenuation coefficient was obtained by the transmission method and results compared with PMMA. The readings were performed by a Radcal ionization chamber, model 10X6-6. The results of attenuation coefficients show that the PLA filament demonstrated a equivalent behavior to PMMA. The PLA plates exhibits an increase in radiation transmission when reduces the infilling, and ABS printing achieved same results for all infills.

Como referenciar
ALMEIDA, J.S.; VILLANI, D.; POTIENS, M.P. A.; WILLEGAIGON, J. Dosimetric characterization of 3D printed for 137Cs gamma rays. Journal of Physics: Conference Series, v. 1826, n. 1, p. 1-6, 2021. DOI: 10.1088/1742-6596/1826/1/012042. Disponível em: http://repositorio.ipen.br/handle/123456789/32496. Acesso em: 26 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções