Performance of three pencil-type ionization chambers (10 cm) in computed tomography standard beams
Carregando...
Data
Data de publicação
2016
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
BRAZILIAN CONGRESS ON METROLOGY, 8th
Resumo
The use of computed tomography (CT) has increased over the years, thus generating a concern about the doses received by patients undergoing this procedure. Therefore, it is necessary to perform routinely beam dosimetry with the use of a pencil-type ionization chamber. This detector is the most utilized in the procedures of quality control tests on this kind of equipment. The objective of this work was to perform some characterization tests in standard CT beams, as the saturation curve, polarity effect, ion collection efficiency and linearity of response, using three ionization chambers, one commercial and two developed at the IPEN.
Como referenciar
CASTRO, MAYSA C. de; XAVIER, MARCOS; CALDAS, LINDA V.E. Performance of three pencil-type ionization chambers (10 cm) in computed tomography standard beams. In: BRAZILIAN CONGRESS ON METROLOGY, 8th, 29 de novembro - 04 de dezembro, 2015, Bento Gonçalves, RS. Proceedings... p. 1-5. Disponível em: http://repositorio.ipen.br/handle/123456789/27218. Acesso em: 30 Dec 2025.