Implementation of a quality assurance program for computerized treatment planning systems

dc.contributor.authorCAMARGO, PRISCILLA R.T.L.pt_BR
dc.contributor.authorRODRIGUES, LAURA N.pt_BR
dc.contributor.authorFURNARI, LAURApt_BR
dc.contributor.authorRUBO, RODRIGO A.pt_BR
dc.coverageInternacionalpt_BR
dc.date.accessioned2014-07-15T13:44:58Zpt_BR
dc.date.accessioned2014-07-30T11:48:10Z
dc.date.available2014-07-15T13:44:58Zpt_BR
dc.date.available2014-07-30T11:48:10Z
dc.date.issued2007pt_BR
dc.description.abstractIn the present investigation, the necessary tests for implementing a quality assurance program for a commercial treatment planning system (TPS), recently installed at Sao Paulo University School of Medicine Clinicas Hospital—Brazil, was established and performed in accordance with the new IAEA publication TRS 430, and with AAPM Task Group 53. The tests recommended by those documents are classified mainly into acceptance, commissioning (dosimetric and nondosimetric), periodic quality assurance, and patient specific quality assurance tests. The recommendations of both IAEA and AAPM documents are being implemented at the hospital for photon beams produced by two linear accelerators. A Farmer ionization chamber was used in a urn:x-wiley:00942405:media:mp2779:mp2779-math-0001 phantom with a dose rate of 320 monitor unit (MU)/min and 50 MU in the case of the dosimetric tests. The acceptance tests verified hardware, network systems integration, data transfer, and software parameters. The results obtained are in good agreement with the specifications of the manufacturer. For the commissioning dosimetric tests, the absolute dose was measured for simple geometries, such as square and rectangular fields, up to more complex geometries such as off-axis hard wedges and for behavior in the build up region. Results were analysed by the use of confidence limit as proposed by Venselaar et al. [Radio Ther. Oncol. 60, 191–201 (2001)]. Criteria of acceptability had been applied also for the comparison between the values of MU calculated manually and MU generated by TPS. The results of the dosimetric tests show that work can be reduced by choosing to perform only those that are more crucial, such as oblique incidence, shaped fields, hard wedges, and buildup region behavior. Staff experience with the implementation of the quality assurance program for a commercial TPS is extremely useful as part of a training program.
dc.format.extent2827-2836pt_BR
dc.identifier.citationCAMARGO, PRISCILLA R.T.L.; RODRIGUES, LAURA N.; FURNARI, LAURA; RUBO, RODRIGO A. Implementation of a quality assurance program for computerized treatment planning systems. <b>Medical Physics</b>, v. 34, n. 7, p. 2827-2836, 2007. DOI: <a href="https://dx.doi.org/10.1118/1.2742779">10.1118/1.2742779</a>. Disponível em: http://repositorio.ipen.br/handle/123456789/5167.
dc.identifier.doi10.1118/1.2742779
dc.identifier.fasciculo7pt_BR
dc.identifier.issn0094-2405pt_BR
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/5167pt_BR
dc.identifier.vol34pt_BR
dc.relation.ispartofMedical Physicspt_BR
dc.rightsclosedAccessen
dc.subjectplanningpt_BR
dc.subjectquality assurancept_BR
dc.subjecttwo-dimensional calculationspt_BR
dc.subjectthree-dimensional calculationspt_BR
dc.subjectmonte carlo methodpt_BR
dc.subjectsimulationpt_BR
dc.subjectdosimetrypt_BR
dc.subjectradiotherapy
dc.subjectcomputerized tomography
dc.titleImplementation of a quality assurance program for computerized treatment planning systemspt_BR
dc.typeArtigo de periódicopt_BR
dspace.entity.typePublication
ipen.autorLAURA NATAL RODRIGUES
ipen.autorPRISCILLA ROBERTA TAVARES LEITE CAMARGO
ipen.codigoautor324
ipen.codigoautor3463
ipen.contributor.ipenauthorLAURA NATAL RODRIGUES
ipen.contributor.ipenauthorPRISCILLA ROBERTA TAVARES LEITE CAMARGO
ipen.date.recebimento08-07pt_BR
ipen.identifier.fi3.198pt_BR
ipen.identifier.ipendoc12724pt_BR
ipen.identifier.iwosWoSpt_BR
ipen.range.fi3.000 - 4.499
ipen.type.genreArtigo
relation.isAuthorOfPublication11143e67-5715-4371-8f83-c77802e0ef1c
relation.isAuthorOfPublicationac01fc3b-f513-40f8-959c-627b2314cb3c
relation.isAuthorOfPublication.latestForDiscoveryac01fc3b-f513-40f8-959c-627b2314cb3c
sigepi.autor.atividadeCAMARGO, PRISCILLA R.T.L.:3463:28:Spt_BR
sigepi.autor.atividadeRODRIGUES, LAURA N.:324:28:Npt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
12724.pdf
Tamanho:
268.5 KB
Formato:
Adobe Portable Document Format

Coleções