Resumo
Specific activity results of 51Cr, 67Ga, 99mTc and 201Tl radioactive solutions, measured in different dose
calibrators located at nuclear medicine services and in a calibrated HPGe gamma spectrometer, were
compared. The HPGe spectrometer was calibrated in a well defined geometry by means of 60Co, 133Ba, 152Eu, 166mHo and 241Am sources, previously standardized in a 4pbg coincidence system. Despite the
observed differences, the results may be accepted within the 10% uncertainty range, established by
Brazilian regulatory standards
Como referenciar
KOSKINAS, MARINA F.; YAMAZAKI, IONE M.; BESSA, ANA C.M.; CALDAS, LINDA V.E.; DIAS, MAURO S. Comparative measurements in dose calibrators and a gamma-ray spectrometer. Applied Radiation and Isotopes, v. 68, n. 4-5, p. 589-591, 2010. DOI: 10.1016/j.apradiso.2009.09.010. Disponível em: http://repositorio.ipen.br/handle/123456789/4726. Acesso em: 30 Dec 2025.