Evaluation of the air kerma rates stability used in diagnostic radiology calibration

Carregando...
Imagem de Miniatura

Data

Data de publicação

2008

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The objective of this work was to perform tests in order to evaluate the stability of the air kerma rates measured with the dosimetric reference system of the calibration laboratory of IPEN (LCI)). The diagnostic radiology reference system consists of a parallel plates ionization chamber with volume of 1cm3 , PTW, model 77334, connected to a PTW electrometer, model UNIDOS, type 10001, and it is used a 14C as check source. This parallel plate ionization chamber can be also used for radiography and fluoroscopy modes measurements during the characterization of the diagnostic radiology beams. The air kerma rates obtained during the calibration measurements were are evaluated from 2005 to 2007.

Como referenciar
FRANCISCATTO, PRISCILA C.; VIVOLO, VITOR; POTIENS, MARIA da P.A. Evaluation of the air kerma rates stability used in diagnostic radiology calibration. In: INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, October 19-24, 2008, Buenos Aires, Argentina. Proceedings... Disponível em: http://repositorio.ipen.br/handle/123456789/15942. Acesso em: 20 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções