Evaluation of the air kerma rates stability used in diagnostic radiology calibration

dc.contributor.authorFRANCISCATTO, PRISCILA C.pt_BR
dc.contributor.authorVIVOLO, VITORpt_BR
dc.contributor.authorPOTIENS, MARIA da P.A.pt_BR
dc.coverageInternacionalpt_BR
dc.creator.eventoINTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATIONpt_BR
dc.date.accessioned2014-11-17T18:02:46Zpt_BR
dc.date.accessioned2014-11-18T18:15:45Zpt_BR
dc.date.accessioned2015-04-02T05:47:20Z
dc.date.available2014-11-17T18:02:46Zpt_BR
dc.date.available2014-11-18T18:15:45Zpt_BR
dc.date.available2015-04-02T05:47:20Z
dc.date.eventoOctober 19-24, 2008pt_BR
dc.description.abstractThe objective of this work was to perform tests in order to evaluate the stability of the air kerma rates measured with the dosimetric reference system of the calibration laboratory of IPEN (LCI)). The diagnostic radiology reference system consists of a parallel plates ionization chamber with volume of 1cm3 , PTW, model 77334, connected to a PTW electrometer, model UNIDOS, type 10001, and it is used a 14C as check source. This parallel plate ionization chamber can be also used for radiography and fluoroscopy modes measurements during the characterization of the diagnostic radiology beams. The air kerma rates obtained during the calibration measurements were are evaluated from 2005 to 2007.
dc.event.siglaIRPA12pt_BR
dc.identifier.citationFRANCISCATTO, PRISCILA C.; VIVOLO, VITOR; POTIENS, MARIA da P.A. Evaluation of the air kerma rates stability used in diagnostic radiology calibration. In: INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, October 19-24, 2008, Buenos Aires, Argentina. <b>Proceedings...</b> Disponível em: http://repositorio.ipen.br/handle/123456789/15942.
dc.identifier.orcidhttps://orcid.org/0000-0002-4049-6720
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/15942pt_BR
dc.local.eventoBuenos Aires, Argentinapt_BR
dc.rightsopenAccesspt_BR
dc.subjectradiologypt_BR
dc.subjectdiagnostic techniquespt_BR
dc.subjectcalibration standardspt_BR
dc.subjectionization chamberspt_BR
dc.subjectkermapt_BR
dc.titleEvaluation of the air kerma rates stability used in diagnostic radiology calibrationpt_BR
dc.typeTexto completo de eventopt_BR
dspace.entity.typePublication
ipen.autorPRISCILA CERUTTI FRANCISCATTO
ipen.autorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.autorVITOR VIVOLO
ipen.codigoautor6403
ipen.codigoautor346
ipen.codigoautor406
ipen.contributor.ipenauthorPRISCILA CERUTTI FRANCISCATTO
ipen.contributor.ipenauthorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.contributor.ipenauthorVITOR VIVOLO
ipen.date.recebimento09-03pt_BR
ipen.event.datapadronizada2008pt_BR
ipen.identifier.ipendoc13393pt_BR
ipen.notas.internasProceedingspt_BR
ipen.type.genreArtigo
relation.isAuthorOfPublicationce9953f2-e814-47e9-89fc-62927f2cb42d
relation.isAuthorOfPublication30f213ae-321d-4d51-8763-0a2585255fb9
relation.isAuthorOfPublication6875184b-48f2-43e7-b6de-e676ee5db15a
relation.isAuthorOfPublication.latestForDiscoveryce9953f2-e814-47e9-89fc-62927f2cb42d
sigepi.autor.atividadeFRANCISCATTO, PRISCILA C.:6403:330:Spt_BR
sigepi.autor.atividadeVIVOLO, VITOR:406:330:Npt_BR
sigepi.autor.atividadePOTIENS, MARIA da P.A.:346:330:Npt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
13393.pdf
Tamanho:
144.93 KB
Formato:
Adobe Portable Document Format

Coleções