Determination of the kinetic parameters of BeO thermoluminescent samples using different methods
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Resumo
The objective of this work was to determine the kinetic parameters of BeO thermoluminescent (TL) samples. The knowledge of the kinetic parameters is important for the general description of the physical characteristics of a thermoluminescent material. Using a linear heating rate of 5Ks-1, the BeO peak temperatures obtained were (474.7 ± 1.2) K and (620.5 ± 1.2) K, respectively, for the first and second TL peaks. The peak shape and the glow curve area measurement methods and the Urbach formula have been used for the trap parameter determination. A kinetic order of about two was determined for the first peak, whereas the second peak followed the first-order kinetics according to the geometrical factor. The energy activation values obtained using different kinds of measurements are in good agreement, including the Kitis method for deconvolution of the glow curve. Furthermore, the results agree with those presented by other authors.
Como referenciar
ALGARVE, F.J.; CALDAS, L.V.E. Determination of the kinetic parameters of BeO thermoluminescent samples using different methods. Brazilian Journal of Radiation Sciences, v. 6, n. 2, p. 1-12, 2018. DOI: 10.15392/bjrs.v6i2.426. Disponível em: http://repositorio.ipen.br/handle/123456789/29935. Acesso em: 24 Mar 2026.